3.3 Error Output
(1)
Error Output
If the vacuum pump of 6-channel degassing unit is in either
status below, an error will be output and the vacuum pump
will stop. At the time of error output, the monitor lamp of
this degassing unit lights in orange.
(a) When the internal pressure of degassing chamber has
not reached 11 kPa at 10 minutes after start of the
vacuum pump drive.
(b) When vacuum level has exceeded 13 kPa.
(c) At occurrence of pressure sensor error
NOTE:
When an error output is issued in the 6-channel
degassing unit, analysis cannot be started from
CDS.
(2)
Release of Error
Error status can be released by turning on power supply to
the Chromaster 5110 pump again.
NOTE:
If the error recurs despite restart of the pump, be
sure to consult your dealer or nearest Hitachi
High-Technologies service representative.