4.2 Periodic Maintenance

4.2.1 Liquid Leak Sensor Check

The Chromaster 5160 pump is equipped with a liquid leak
sensor for detecting liquid leakage. Note, however, that liquid
leakage may not be detected correctly due to sensitivity of the
leak sensor. To prevent this, perform periodic operation check
for the liquid leak sensor as instructed below.
(1)
Operation check
Put 0.5 mL of water on the liquid leak sensor. And confirm
the leak error.
Put 0.5 mL of water
Fig. 4-2
Leak Sensor Position
NOTES: 1.
When a solvent is used, turn on the liquid leak
sensor function.
2.
The liquid leak sensor may not detect a drop of
leakage or high-volatile solvents. Always check
the leakage even while the liquid leak sensor is
turned ON.
3.
Visual confirmation of a liquid leak is
recommended as a daily check.
4.
A periodic check for the liquid leak sensor is
required also.